Return to site

PXI Mobile Phones & Portable Devices Driver Download For Windows 10

broken image


Multiple-Device Synchronization Using USB or PXI Devices

  1. Pxi Mobile Phones & Portable Devices Driver Download For Windows 10 Hp
  2. Pxi Mobile Phones & Portable Devices Driver Download For Windows 10 Windows 7

The PXI‑8840 includes two 10/100/1000BASE‑TX (Gigabit) Ethernet ports, two USB 3.0 ports, four USB 2.0 ports, as well as an integrated hard drive, serial port, and other peripheral I/O. The PCI eXtensions for Instrumentation (PXI) modular instrumentation architecture delivers a rugged, PC-based, and high-performance measurement and automation system. With PXI you benefit from the low cost, performance, and flexibility of the latest PC technology and the benefits of an open industry standard. While top-end phones like the iPhone X are pushing prices ever higher, at the bottom end smartphones are cheaper than they've ever been, and the Alcatel Pixi 3 is a prime example. Costing roughly £30 SIM free or from around £7 per month on contract with no upfront cost this is ten times cheaper than even most mid-range phones, yet it runs.

You can synchronize multiple devices in a DataAcquisition using a shared scan clock and shared start trigger. You can synchronize devices using either PFI or RTSI lines.

Requirement

You must register your RTSI cable using the National Instruments™ Measurement & Automation Explorer.

Acquire Synchronized Data Using USB Devices

This example shows how to acquire synchronized voltage data from multiple devices using a shared start trigger and a shared scan clock. Analog input channels on all three devices are connected to the same function generator.

Pxi Mobile Phones & Portable Devices Driver Download For Windows 10 Hp

PXI Mobile Phones & Portable Devices Driver Download For Windows 10

Create a DataAcquisition and add one voltage input channel from each device:

  • NI USB-6211 with device ID Dev1

  • NI USB 6218 with device ID Dev2

  • NI USB 6255 with device ID Dev3

Choose terminal PFI4 on Dev1 as the start trigger source. Connect the trigger source to the destination terminals PFI0 on Dev2 and PFI0 on Dev3.

Choose terminal PFI5 on Dev1 as the scan clock source. Connect it to destination terminals PFI1 on Dev2, and PFI1 on Dev3.

Acquire data and assign it to dataIn.

Plot the data.

All channels are connected to the same function generator, so the plot displays overlapping signals, indicating synchronization.

Pxi Mobile Phones & Portable Devices Driver Download For Windows 10 Windows 7

Synchronize Counter Outputs from Multiple Devices

PXI Mobile Phones & Portable Devices Driver Download For Windows 10

This example shows how to synchronize the start trigger of counter output operations from two channels on different devices.

This example uses two USB or PCI devices, but could be modified for channels across CompactDAQ or PXI chassis. If you have counter output CompactDAQ modules in the same chassis, it is not necessary to call addtrigger; but it is required for multiple modules in the same PXI chassis.

Synchronize DSA PXI Devices Using AutoSyncDSA

This example shows how to acquire synchronized data from two Dynamic Signal Analyzer (DSA) PXI devices, NI PXI-4462 and NI PXI-4461, using the AutoSyncDSA property.

Create a DataAcquisition and add one voltage analog input channel from each of the two PXI devices

Acquire data in the foreground without synchronizing the channels:

The data returned is not synchronized.

Synchronize the two channels using the AutoSyncDSA property:

Acquire data in the foreground and plot it:

The data is now synchronized.

Acquire Synchronized Data Using PXI Devices

This example shows how to acquire voltage data from two PXI devices on the same chassis, using a shared start trigger to synchronize operations within your DataAcquisition. PXI devices have a shared reference clock that automatically synchronizes scan clocking. You need to add only start trigger connections to synchronize operations in your DataAcquisition with PXI devices. Analog input channels on all devices are connected to the same function generator.

Create a DataAcquisition and add one voltage input channel from each NI-PXI 4461 device with IDs PXI1Slot2 and PXI1Slot3.

Add a start trigger connection to terminal PXI_Trig0 on PXI1Slot2 and connect it to terminal PXI_Trig0 on PXI1Slot3. PXI cards are connected through the chassis backplane, so you do not have to wire them physically.

Acquire data and assign it to dataIn.

Plot the data.

PXI
Pxi mobile phones & portable devices driver download for windows 10 windows 7

Create a DataAcquisition and add one voltage input channel from each device:

  • NI USB-6211 with device ID Dev1

  • NI USB 6218 with device ID Dev2

  • NI USB 6255 with device ID Dev3

Choose terminal PFI4 on Dev1 as the start trigger source. Connect the trigger source to the destination terminals PFI0 on Dev2 and PFI0 on Dev3.

Choose terminal PFI5 on Dev1 as the scan clock source. Connect it to destination terminals PFI1 on Dev2, and PFI1 on Dev3.

Acquire data and assign it to dataIn.

Plot the data.

All channels are connected to the same function generator, so the plot displays overlapping signals, indicating synchronization.

Pxi Mobile Phones & Portable Devices Driver Download For Windows 10 Windows 7

Synchronize Counter Outputs from Multiple Devices

This example shows how to synchronize the start trigger of counter output operations from two channels on different devices.

This example uses two USB or PCI devices, but could be modified for channels across CompactDAQ or PXI chassis. If you have counter output CompactDAQ modules in the same chassis, it is not necessary to call addtrigger; but it is required for multiple modules in the same PXI chassis.

Synchronize DSA PXI Devices Using AutoSyncDSA

This example shows how to acquire synchronized data from two Dynamic Signal Analyzer (DSA) PXI devices, NI PXI-4462 and NI PXI-4461, using the AutoSyncDSA property.

Create a DataAcquisition and add one voltage analog input channel from each of the two PXI devices

Acquire data in the foreground without synchronizing the channels:

The data returned is not synchronized.

Synchronize the two channels using the AutoSyncDSA property:

Acquire data in the foreground and plot it:

The data is now synchronized.

Acquire Synchronized Data Using PXI Devices

This example shows how to acquire voltage data from two PXI devices on the same chassis, using a shared start trigger to synchronize operations within your DataAcquisition. PXI devices have a shared reference clock that automatically synchronizes scan clocking. You need to add only start trigger connections to synchronize operations in your DataAcquisition with PXI devices. Analog input channels on all devices are connected to the same function generator.

Create a DataAcquisition and add one voltage input channel from each NI-PXI 4461 device with IDs PXI1Slot2 and PXI1Slot3.

Add a start trigger connection to terminal PXI_Trig0 on PXI1Slot2 and connect it to terminal PXI_Trig0 on PXI1Slot3. PXI cards are connected through the chassis backplane, so you do not have to wire them physically.

Acquire data and assign it to dataIn.

Plot the data.

All channels are connected to the same function generator and have a shared reference clock. The signals overlap, indicating synchronization.

Related Topics





broken image